  • 高温2.png
  • 高温1.png
  • 高温3.png

number

High temperature test chamber


Retail price

Market price


High temperature test chamber

key word:humidity

Consultation hotline:

weight

Quantity
-
+

inventory

隐藏域元素占位

  • Advisory Details
  • Product name

    High temperature test chamber

    Model

    YB91-200N YB216-200N YB512-200N YB1000-200N YB91-300N YB216-300N YB512-300N YB1000-300N

    Interior volume (L)

    91 216 512 1000 91 216 512 1000

    Inside dimension

    (mm)

    W

    450 600 800 1000 450 600 800 1000

    H

    450 600 800 1000 450 600 800 1000

    D

    450 600 800 1000 450 600 800 1000

    Outside dimension (mm)

    W

    1060 1210 1500 1730 1060 1210 1500 1730

    H

    820 970 1210 1480 820 970 1210 1480

    D

    635 785 1065 1275 635 785 1065 1275
    Temperature range (℃) (ambient temperature+20~200) ℃ (ambient temperature+20~300) ℃
    Temperature fluctuation

    ≤100℃

    ±0.2℃ ±0.3℃ ±0.3℃ ±0.3℃
    ≤200℃ ±0.2℃ ±0.4℃ ±0.3℃ ±0.4℃

    ≤300℃

    ±0.4℃ ±0.6℃
    Temperature deviation

    ≤100℃

    ±1.5℃ ±2℃ ±1.5℃ ±2℃
    ≤200℃ ±2.0℃

    ≤300℃

    ±3.0℃ ±3.0℃
    Heating time (min) (ambient temperature+20~200) ℃ (ambient temperature+20~300) ℃
    ≤40/min ≤60/min ≤70/min

    Temperature regulation method

    Forced air circulation heat exchange method
    Usage environment Temperature: (5~36) ℃ Humidity: (30~85)% RH Pressure: (86~106) kPa

    Material Science

    Inner box

    Stainless steel plate SUS304

    Outer box

    Double sided galvanized sheet, surface treated with electrostatic spraying treatment

    Heat preservation

    Fiber high-temperature cotton

    Actuator

    Fan

    Extended shaft centrifugal fan

    Damper

    Manually adjust ventilation and circulation

    Heater

    Imported nickel chromium alloy electric heating wire

    Controller system

    Display

    7.0",800x480 resolution, mixed output PLC

    Operation mode

    Fixed value running, program running

    Setting method

    English, key input

    Display resolution

    Temperature: 0.1 ℃, Time: min

    Sensor

    PT100 (platinum resistance)

    Control mode

    Anti integral saturation P1D+self-tuning

    Communication function

    (Optional function) RS485, convenient for remote and local communication monitoring, (requires monitoring software, RS485/RS232 converter, requires one COM and one USB port for PC), can connect up to 16 units at the same time, with a maximum cable length of 700m

    Supplementary functions

    Fault alarm function, upper and lower limit protection function

    Applicable power supply (V)

    AC voltage 380V ± 38V frequency 50Hz ± 0.5Hz three-phase four wire+grounding protection

    Maximum current (A)

    3 4 7.6 9.8 4.1 5.8 9.8 14.5

    Power (KW)

    2 2.7 5 6.5 2.7 3.8 6.5 9.5
    Weight (kg) 95 130 240 430 95 130 240 430

    Standard configuration

    1 on each side of the cable port (ф50mm) and 2 sets of sample shelf

    Safety device

    Leakage switch, motor, compressor, phase sequence, over temperature, dry burning and other protection functions

The previous

next

资料下载

Request a quotation

*
*
*
*
*
Submit